国产精品亚洲一区二区在线|18禁av一区二区三区|免费成人午夜福利电影79|97精品人妻麻豆一区二区

新聞中心您現(xiàn)在的位置:首頁(yè) > 新聞中心 > WATER CHEMISTRY超純冷卻單元323A2419
WATER CHEMISTRY超純冷卻單元323A2419
更新時(shí)間:2017-08-11   點(diǎn)擊次數(shù):1588次

WATER CHEMISTRY(GE)超純冷卻單元323A2419

 

 

 

 

 

 

 

 

 

 

RAINMAKER Ultra Pure Coolant Purge Pod

 

Specifications below:

 

Deionizer Specification

 

1. General

The application is to maintain the resistance of a closed loop coolant system of 80 gallons at anominal 7 megohm cm. The coolant is either pure water or a mixture of water and laboratory grade Propylene Glycol. The parallel loop flow rate through the pod is 0.7 gpm and the inlet is at 50 psi. The maximum coolant temperature is 155

2. Tank: Similar in size to RT-818 by Park International Corp.

Height: 25 inches maximum with fittings.

Diameter: 8.8 ± 0.1” at the base. Capacity: .45 to .47 cubic feet.

Material: Fiberglass/ester laminate, thermoplastic liner.

Weight: 5 pounds empty, 31 pounds dry, 48 pounds wet.

Pressure rating: 75 psi at 155.

 

3. Mixed be resin of suitable quality

Volume: .45 cubic feet or 100% of the available.

 

4. None (no carbon filter media).

5. Connections (identify inlet and outlet)

Inlet: Colder Products Company part number HFCD 108-35

Outlet: Colder Products Company part number HFCD 248-35

 

6. Shipping container

Domestic: The shipping container must be suitable for common modes of postal or commercial delivery and should be clearly marked with this drawing number and applicable part number. Safety data sheets for the resin in this canister should be placed in the shipping container. This canister is tobe marked with this drawing number and part number.

Export: Water Chemistry, Inc. provides delivery in the United States only; however, a special hard-body shipper is available upon request. Shipping container will conform to the regulations of thereceiving country.

、

有關(guān)美國(guó)WATER CHEMISTRY產(chǎn)品更詳細(xì)的信息或者購(gòu)買(mǎi)美國(guó)WATER CHEMISTRY相關(guān)產(chǎn)品,請(qǐng)愛(ài)澤工業(yè)。